- Machinery Parts Processing Services[6]
- Cutting and Forming Tools[1]
- Cleaning Equipment[8]
- Air Cleaning Equipment Parts[4]
- Filters[1]
- Generators[10]
- Spectrum Analyzers[1]
- Machinery Agents[1]
- Relays[79]
- Other Physical Measuring Instruments[1]
- Testing Equipment[10]
- Other Woodworking Machinery[12]
- Measuring & Analysing Instrument Stocks[4]
- Other Electronic Measuring Instruments[7]
- Other Electrical Instruments[6]
- Other Analysis Instruments[9]
- Weighing Scales[7]
- Ultrasonic[5]
- PH Meters[4]
- Laboratory Thermostatic Devices[7]
- Sandwich Panels[7]
- Densitometers[2]
- Moisture Meters[2]
- Other Optics Instruments[2]
- Spectrometers[1]
- Other Measuring & Analysing Instruments[3]
- Mixing Equipment[4]
- Drying Equipment[3]
- Sterilization Equipments[7]
- Concentration Meters[2]
- Counters[7]
- Laboratory Furniture[7]
- Packaging Bags[7]
- Air Purifiers[4]
- Gas Disposal[1]
- Vacuum Cleaners[1]
- Steam Cleaners[1]
- Other Air Cleaning Equipment[1]
- Contact Person : Mr. Tu Victor
- Company Name : Shanghai Selon Scientific Instrument Co., Ltd.
- Tel : 86-21-35100022
- Fax : 86-21-65416501
- Address : Shanghai,Shanghai,29F, Building 2nd, Lane 388, Kunming Road
- Country/Region : China
- Zip : 200082
Gas Chromatography
Flexible Sample Introduction System
Up to three injectors may be installed and operated simultaneously with independent temperature control. Wide range of injectors is available to be chosen from for various applications.
Sample injector for single/dual packed columns
- Several alternative sample introduction methods, including on-column sample introduction and six-way valve injection;
- Adapter for injection to capillary column (0.53 mm i.d.).
Sample injector for capillary column
- Special sample injector with septum purge and back-pressure valve splitting
- Special programmable split/splitless injector with solenoid valve, septum purge, back-pressure valve splitting, and standalone carrier gas and make-up gas flow
- Modified dual-packed-column sample injector for injection to capillary column by employing some connectors and back-pressure valve splitting.
Advanced Pneumatics
1. Carrier gas control with constant pressure valve and constant flow valve;
2. Air and hydrogen control with constant pressure valve and needle valve;
3. Flow easily adjustable through constant flow valve and needle valve with digital display;
4. High precision, reproducibility and reliability.
Sensitive and Reliable Detector
Up to three detectors (FID, TCD, ECD, etc.) may be installed and operated simultaneously. Simultaneous multi-detection with detectors in series or in parallel meets the various analysis requirements.
Advanced Microcomputer Control System
With advanced control techniques, microcomputer-based temperature control system provides superior performance.
- High temperature precision (± 0.5°C), high reliability, and high resistance to disturbance;
- Five independent heated zones, with maximum operating temperature of 400°C;
- Temperature-limit setting and overheat protection.
Instrument control is accomplished through the keypad, and the instrument offers outstanding capabilities.
- Self-diagnosis;
- Power-off protection;
- Ten methods saving and reloading;
- Two controls of external events;
- Timing on/off of temperature control;
- Timing program, detector status, measurement range, polarity and current setting and display;
- Temperature setting values, actual values, retention time and analysis time display.
High-performance spacious column oven
1. Spacious column oven accommodating capillary column and dual packed columns and convenient for column installation;
2. Fast heat-up and rapid cool-down column oven with automated back-door opening (300°C to 50°C in 7 min);
3. Available for both near-ambient and above ambient temperature control;
4. Supports 5 oven ramps with 6 plateaus.
Specification:
GC112A-1 | Sample injector with special-purpose capillary column, having the blowing-up function of diaphragm and the shunting regulation with back pressure valves+ make-up gas adjustment+ FID |
GC112A-2 | Sample injector with Double Packed Column + TCD |
GC112A-3 | Sample injector with Packed Column + FID |
GC112A-4 | Sample injector with Double Packed Column + The (annex or appendix or enclosure) with special-purpose capillary Column+ back pressure valves+Double FID |
GC112A-5 | Sample injector with Double Packed Column + Special-purpose capillary column having the blowing-up function of diaphragm and the shunting regulation with back pressure valves + Double FID |
GC112A-6 | Sample injector with Double Packed Column + Sample injector with special-purpose capillary column, having carrier gas control, the blowing-up function of diaphragm and the shunting regulation with back pressure valves+ make-up gas adjustment + Double FID |
Instrument technical index and operational requirement
Column oven temperature index
Temperature range | 7°C ~ 400°C above room temperature, (increment 1°C) |
Temperature control accuracy | better than ±0.1°C at 200°C |
Temperature program ramps | 5 ramps in total |
Program setting | 0.1°C~40°C/min, (increment 0.1°C) at 200°C |
Total time of total program | Up to 655 minutes, (increment 0.1°C) |
Injector and FID temperature index
Temperature range | 7°C~400°C above room temperature, (increment 1°C) |
Temperature control accuracy | better than ±0.1°C at 200°C |
Hydrogen flame ionization detector (FID)
Minimum detectable quantity | Dt ≤1×10-11 g/s (n-hexadecane) |
Fine road-test | Dt ≤8×10-12 g/s (n-hexadecane) |
Drift | ≤6×10-13 A/h |
Linear dynamic range | ≥106 |
Maximum temperature | 400°C |
Thermal conductivity detector (TCD)
Sensitivity | S≥2500mv.ml/mg (n-hexadecane) |
Tiptop Sensitivity | S≥10000mv.ml/mg (with amplifying circuit) |
Noise | ≤20 μ V |
Baseline drift | ≤30 μ V /h |
Linear range | ≥104 |
Maximum temperature | 400°C |
Instrument operational requirement
Power source | 220V~±22V 50Hz ± 0.5Hz |
Power | ≤2000W |
Ambient temperature | +5°C ~ +35°C |
Relative humidity | ≤85% |
Gas Chromatography